Bird-technologies 5019D-Manual Bedienungsanleitung Seite 18

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 59
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 17
6
If calibration fails, “Cal Fail” will be displayed.
a. Press a key to return to normal operation,
b. Ensure that the WPS is properly connected, and the
RF is off.
c. Rezero. Go back to Step 3.
5000-XT
1. Make sure the sensor has reached a stable operating
temperature.
2. Make sure no RF power is applied to the sensor.
3. Press and hold “0” for two seconds to begin zero cali-
bration.
WARNING
Do not interrupt the calibration.
Note: “Zero/Cal” will be displayed and calibration
will begin.
Note: Zeroing the sensor takes 60 seconds. The
bar graph will display calibration progress.
Note: When complete, “PASS” should be displayed.
4. Press Enter to return to normal operation when zeroing is
complete.
Note: If “FAIL” is displayed, make sure no RF power
is applied to the sensor and perform the procedure
again.
Seitenansicht 17
1 2 ... 13 14 15 16 17 18 19 20 21 22 23 ... 58 59

Kommentare zu diesen Handbüchern

Keine Kommentare